EXTERNAL CONTROL
175
CR Mode
Proportional Control:
Input conductance =
rated conductance × (external resistance/10).
Inverse Control:
Input conductance =
rated conductance × (1 - external resistance/10).
Input
Conductance
External
Resistance
Rated
Conductance
0
Ω
10k
Ω
Proportional control
Inverse control
CV Mode
Proportional Control:
Input voltage = rated voltage × (external
resistance/10).
Inverse Control:
Input voltage = rated voltage × (1 - external
resistance/10).
Input
Voltage
External
Resistance
Rated
Voltage
0
Ω
10k
Ω
Proportional control
Inverse control