OPERATION
83
Display
8.653
V
50.006
mA
0.4327
W
CV B Value
80.000
CV A Value
80.000
LOAD
RS232
A Value
Fine
01/Oct/2017
V
V
OCP Alarm
Please Press Enter
To Clear Alarm
OCP
OCP indicator
Alarm message
when OCP is
set to Load Off
Mode
CV
I Range
L87.5mA
V Range
L 80V
Response
Fast
Configure
OPP
Description
For OPP, the PEL-3000(H) can be configured to
either limit the power or turn off the load.
The OPP levels can be set to 10% higher than the
rating power.
Operation
1.
Press
Main
>
Configure[F5]
>
Protection[F1]
and set the
OPP Level
and
OPP Setting
.
Range:
OPP Level: Rating power + 10%
OPP Setting: LIMIT, Load Off
Alarm
When
OPP Setting
is configured to
Load Off
, a
message will be displayed on the screen when
OPP is tripped. The Enter key must be pressed
to clear the alarm message.
When configured to
LIMIT
, OPP will be
displayed on the screen when the OPP is
tripped and the power will be limited to the
OPP Level
setting.
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...