PEL-3000(H) User Manual
190
Parallel Capacity, PEL-3211H
Model
No. of
Units
V
I
Total Sink Current
PEL-3111H+
PEL-3211H
Total Power
PEL-3111H +
PEL-3211H
PEL-3111H:
Master
x 1
800V 52.5A N/A
N/A
PEL-3211H:
Slave
Boosters
x 1
800V 105A 157.5A
3150W
x 2
800V 210A 262.5A
5250W
x 3
800V 315A 367.5A
7350W
x 4
800V 420A 472.5A
9450W
Note
The PEL-3211H booster packs do not have a control
panel. They can only be used as slaves with a single
PEL-3111H in parallel.
Connection
Description
The J1 and J2 connectors are used for control
during parallel operation. Up to 5 units can be
used in parallel.
Note
Only the rear panel terminals can be used for parallel
operation, the front panel terminals have a lower
current rating and thus should not be used for parallel
operation.
Содержание PEL-3000 Series
Страница 1: ...DC Electronic Load PEL 3000 H Series USER MANUAL VERSION 2 00 ISO 9001 CERTIFIED MANUFACTURER ...
Страница 24: ...PEL 3000 H User Manual 22 PEL 3111H PEL 3211 Booster Pack PEL 3211H Booster Pack ...
Страница 29: ...GETTING STARTED 27 GRA 413 JIS standard 149 532 5 482 6 GRA 414 E EIA standard 132 382 5 477 6 ...
Страница 41: ...GETTING STARTED 39 Fasteners 2 ...
Страница 207: ...REMOTE CONTROL 205 6 7 8 9 The function check is complete ...
Страница 209: ...FAQ 207 For more information contact your local dealer or GWInstek at www gwinstek com marketing goodwill com ...
Страница 229: ...APPENDIX 227 Operating Area PEL 3021 ...
Страница 230: ...PEL 3000 H User Manual 228 PEL 3041 ...
Страница 231: ...APPENDIX 229 PEL 3111 ...
Страница 232: ...PEL 3000 H User Manual 230 PEL 3211 Booster Pack PEL 3021H ...
Страница 233: ...APPENDIX 231 PEL 3041H ...
Страница 234: ...PEL 3000 H User Manual 232 PEL 3111H PEL 3211H Booster Pack ...
Страница 235: ...APPENDIX 233 ...