Section 6: Troubleshooting
This section of the manual provides information on error messages,
troubleshooting and servicing the Agrifeed 5500 indicator.
6.1 Operational Mode Error Messages
Message Description
02 UnderLoad!
Input signal less than negative full scale. If this is due to
excessive loading, reduce the load. Otherwise check the load
cell connections. Verify that the scale capacity in section 3.3.2
is correct.
03 Over-Load!
Input signal is greater than positive full scale. Use same check
as for underload.
04 # > Dsply
Number to be displayed will not fit within 6 digits. This will
not normally occur for the Gross, Net or Tare Weights. Either
clear the totals or settle for only being able to transmit the
totals.
05 Zero> Max.!
An attempt was made to zero out more than allowed. The
allowed value is 10% of scale capacity. Use the
[TARE]
key
for subtracting off container weights or if large dead-load is
always to be present, apply this dead-load during the No Load?
prompt during calibration to permanently eliminate the offset.
06 Tare>F.S.!
Tare entry was greater than full scale. Most likely the entered
tare value was incorrect.
07 Tare < 0 !
Negative tare attempted, but not allowed. For auto-tares, the
GROSS Weight must be greater than zero unless.
08 CheckConn.
This message is displayed if the signal into the A/D is +/- 2
times the Full Scale signal. This is effectively taken into
consideration when the information sent to the micro processor
from the A/D is +/- twice the allowable F.S. reading. For
example, Scale Capacity = 100. Error message will be
displayed at +/- 208 taking into consideration the 4% overload.