32
00
New
Zero?
Calibration Mode
Figure 24: Calibration Mode
There are six Calibration Mode selections as described below:
"New
Zero?"
Most common calibration procedure used to
establish a new zero (no load) and span (test
load) calibration reference.
"Last
Zero?"
Allows a span re-calibration without removing
the test load.
"Temp
Zero?"
Allows calibration without removing the
current gross weight. The zero reference
determined during the last calibration is
maintained.
"Only
Zero?"
Establishes a new zero reference without
affecting span. Useful when changing the
scale's dead load, for example adding safety
rails to a scale platform.
"Cal
Reset"
Adjusts the zero and gain factors of the A/D
amplifier to default values for maximum
sensitivity.
"Known
LCOut"
Method of calibrating without the use of test
weights. The loadcell mV/V value and full
scale capacity must be known.
Press
[NET/GROSS]
to scroll through and select a calibration method.
Press
[ENTER]
to begin the calibration method selected and refer to the
appropriate section for calibration instructions:
New Zero?
See section 3.3.7.1
Last Zero?
See section 3.3.7.2
Temp Zero?
See section 3.3.7.3
Only Zero?
See section 3.3.7.4
Cal Reset
See section 3.3.7.5
Known LCOut
See section 3.3.7.6