3.3.7 Calibrate
Mode
The Calibration Mode is used to calibrate the Agrifeed 5500 scale system
to a known weight value, thus establishing an accurate weight indication
throughout the weighing range. Access the Calibration selection as
described in section 3.3. From the Calibration Setup Menu Selection,
press
[YES]
to access the calibration routines (see Figure 24).
SETUP MENU SELECTION
J
CALIBRATION
I
[YES]=Access [NO]=Next [HELP]=Exit
Figure 23: Setup Menu Selection - Calibration
Notes on Calibration:
Press
[CLR]
at any point in the calibration routine to move back one
step.
Pressing
[CLR]
at the
New Zero?
prompt will exit the calibration
mode.
If a calibration weight is less than 5% of capacity, or if there was a
large change in the calibration, the display prompts
ReCal Req'd
suggesting that the calibration process be repeated. Press
[YES]
to
repeat the calibration procedure, or press
[NO]
accept the
calibration and obtain the
CAL OK?
prompt.
If a calibration weight exceeds full scale by +4% or is less than 0.1%
of full scale, an error message is displayed
A calibration weight may be applied before or after the calibration
weight is entered. The display will prompt you to
Keyin CalWT
(key
in calibration weight) or
Add CalWT
(add calibration weight) at the
appropriate time.