6.3.6 BREATH CYCLE TIME
SVS-5033_XX
This is the length of the breath i.e. the total
cycle time of the mandatory breath in SIMV
(inspiration, pause plus expiration).
Note: The touchpad Breath cycle time is
not shown when an SIMV mode is selected
and inspiration time is configured. Refer to
heading I:E ratio / Inspiration times.
6.3.7 VOLUME SETTING
Depending on the ventilator configuration the
inspiratory volume can be set as:
Minute Volume or,
Tidal Volume
Note: The configuration is done by a service
technician with a service card.
6.3.8 CONTROLLED / SUPPORTED
PRESSURE LEVEL
PC (Pressure Control level) above PEEP is the
set inspiratory pressure level for each
mandatory breath in Pressure Control and
SIMV (PC) + PS, and also for backup
ventilation in Pressure Support.
PS (Pressure Support level) above PEEP is the
set inspiratory pressure support level for
triggered breaths in Pressure Support, SIMV
modes and Bi-Vent.
6.3.9 O
2
CONCENTRATION
The setting range for the gas mixer is 21% O
2
to 100% O
2
. The alarm limits are automatically
set at approximately 5% O
2
above or below
the set concentration value. The alarm is
delayed 40 seconds after changing the O
2
concentration setting. There is also an
absolute minimum alarm limit of 18% O
2
which
is independent of operating settings.
6.3.10 RESPIRATORY RATE / SIMV RATE
Respiratory rate is the number of controlled
mandatory breaths per minute in controlled
modes excluding SIMV. The respiratory rate
is also used for calculation of tidal volume if
the ventilator is configured for Minute volume
setting. SIMV rate is the number of controlled
mandatory breaths in SIMV modes.
76
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 6 |
Ventilation, modes and functions
|
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...