6.9.6 SIMV - IN DETAIL
3
4
1
2
SVX-9027_EN
SIMV
Breath cycle
time
V
P
Spont. period
SIMV
Breath cycle
time
90%
time
time
1.
This combined control and pressure
support/spontaneous function allows for
preset mandatory breaths synchronized
with the patient's breathing.
2.
If there is no trigger attempt within a time
window equal to 90% of the set Breath
cycle time, a mandatory breath is
delivered. (The Breath cycle time is the
total time for one mandatory breath.)
3.
The mandatory breath is defined by the
basic settings (mode of ventilation, breath
cycle time, respiratory pattern and
volumes/pressures).
4.
The spontaneous/pressure supported
breaths are defined by the setting for
Pressure Support.
100
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 6 |
Ventilation, modes and functions
|
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...