8.3 NEBULIZERS
8.3.1 GENERAL
The nebulizer is intended for administering
drugs to patients requiring mechanical
ventilation or positive pressure breathing
assistance via an endotracheal tube or face
mask/prongs.
The nebulizer operates continuously
regardless of ventilation mode setting. No
extra gas volume is added to the inspiratory
minute volume and ventilator settings and
values are not affected
The SERVO-s Ventilator System must not be
used with jet nebulizers.
Please refer to the manufacturer's operating
manual for instructions on use.
8.3.2 AERONEB NEBULIZER SYSTEM
See Aeroneb Nebulizer System's own User
Manual for instructions.
WARNINGS!
Disconnect the Servo Humidifier/HME
during nebulization; otherwise the
humidifier may become blocked or the
drug may be trapped in the humidifier
Do not use the nebulizer without a filter
e.g. Servo Duo Guard, connected to the
expiratory inlet of the ventilator.
During nebulization, carefully monitor the
airway pressure. Increased airway
pressure could result from a clogged
filter. Replace the filter if the expiratory
resistance increases or after maximum
usage time according to filter
specification, whichever comes first.
CAUTION: If the Aeroneb nebulizer is used
with active humidification, then the particle
size of the medication may be affected.
CAUTION: The nebulizer must not be left
unattended when connected to a patient.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
129
|
Optional Accessories
| 8 |
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...