7.2.3 PRE-SILENCING ALARMS
To silence most alarms (active and inactive)
for two minutes, press and hold the Audio
Pause fixed key for more than two seconds.
This action has the following results:
1.
All alarms, active and inactive, are
silenced for two minutes from the time the
key was pressed.
2.
A double crossed bell symbol along with
the time remaining in the silent period are
displayed in the message area.
3.
Pressing the Audio Pause key again for
less than two seconds will now restart the
silent period for two additional minutes
from when the user last pressed the key.
4.
Latched alarms are reset if the alarm
condition has ceased.
SVX-5098_EN
When the alarms are pre-silenced, pressing
and holding the Audio Pause key again for
more than 2 seconds will reactivate all audible
alarms.
124
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 7 |
Alarms
|
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...