4.10 SUCTIONING
4.10.1 SUCTION SUPPORT
The Suction Support function makes it
possible to automatically inhibit the ventilator
from cycling during a tracheal suction
procedure without activating alarms.
Suction Support includes:
pre-oxygenation/preparation phase
disconnect phase
post-oxygenation phase
WARNINGS!
Suction Support is not intended to be
used together with closed-suction
systems.
The minimum PEEP level during suction
support is 3 cmH
2
O. The ventilator will
adjust to the minimum level if the PEEP
level is below 3 cmH
2
O in order to detect
disconnection of the patient.
Important: Alarms are turned off during
the disconnect phase for a maximum of 60
seconds. If the patient has not been
reconnected within 60 seconds, alarms are
activated.
Notes:
Suction Support is not available in NIV
mode or when the O
2
Breaths function is
activated.
During the disconnect phase in Suction
Support, the nebulizer is temporarily
paused.
When only one gas is connected, an
elevated oxygen level cannot be set
during the preparation phase. In this
case, the post-oxygenation phase will be
skipped.
4.10.2 PRE-OXYGENATION/PREPARATION
PHASE
SVS-9000 XX
To enter the pre-oxygenation/preparation
phase:
1.
Press the Quick access fixed key.
2.
Press the Suction Support touchpad.
3.
Set the desired pre-oxygenation value by
turning the Main Rotary Dial and press
Accept.
56
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 4 |
Operation overview
|
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...