The Check tubing alarm is turned off: the
maximum duration of the preparation phase
is 120 seconds. After 120 seconds, the system
automatically returns to ventilation using the
previous oxygen setting.
Note: The Cancel touchpad will close the
Suction Support program.
4.10.3 DISCONNECT PHASE
The system automatically enters the
disconnect phase when the patient is
disconnected during the
pre-oxygenation/preparation phase.
During the disconnect phase the following
alarms are turned off for up to 60 seconds:
Apnea
Minute volume
Respiratory rate
PEEP
When the patient is reconnected, the system
automatically enters the post-oxygenation
phase and restarts ventilation.
It is also possible to restart ventilation
manually by pressing the Start ventilation
touchpad.
4.10.4 POST-OXYGENATION PHASE
After reconnection, the ventilator will deliver
the same oxygen concentration as in the
preparation phase for 60 seconds.
After 60 seconds the system automatically
returns to ventilation using the previous
oxygen setting.
4.10.5 CLOSED-SUCTION SYSTEMS
If a closed-suction system is used, the Suction
Support, Insp. hold and Exp. hold functions
should
not
be used. Pressure-based modes
(such as Pressure Control, Pressure Support,
Bi-Vent, or SIMV (PC) + PS) should be used.
Settings should be adjusted to levels suitable
for the patient. Hospital guidelines for
suctioning should be followed.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
57
|
Operation overview
| 4 |
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...