6.10 BI-VENT
6.10.1 FUNCTIONAL DESCRIPTION
S
V
S
-5056_XX
Bi-Vent is pressure controlled breathing that
allows the patient the opportunity of
unrestricted spontaneous breathing. Two
pressure levels are set together with the
individually set duration of each level.
Spontaneous breathing efforts can be assisted
by pressure support
The following parameters are set:
1.
Pressure high (PHigh) for the higher
pressure level (cmH
2
O)
2.
PEEP for the lower pressure level (cmH
2
O)
3.
Oxygen concentration (%)
4.
Time at the higher pressure (THigh) level
(s)
5.
Time at the lower pressure (TPEEP) level
(s)
6.
Inspiratory rise time (s)
7.
Trigg. Flow / Trigg. Pressure
8.
Inspiratory Cycle-off (%)
9.
PS (Pressure Support level) above PHigh
(cmH
2
O)
10. PS (Pressure Support level) above PEEP
(cmH
2
O)
S
V
S
-5057_XX
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
103
|
Ventilation, modes and functions
| 6 |
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...