13.2 SPECIAL FUNCTION KEYS
The ventilator will initiate a new breath cycle according to the current ventilator
settings.
Start breaths
This function allows 100% oxygen to be given for 1 minute. After this time the
oxygen concentration will return to the pre-set value. The oxygen breaths can be
interrupted by pressing the O
2
breaths fixed key during the 1 minute interval.
Note: If O
2
breaths is activated during the pre- or post-oxygenation phase in
Suction Support the procedure will be discontinued.
O
2
breaths
Expiratory hold is activated by manually pressing the Exp. hold key. The maximum
time is 30 seconds. The inspiratory and expiratory valves close after expiration.
This function can provide an exact measurement of the end expiratory lung
pressure. It can be used for static compliance measurement and to determine the
total PEEP.
Exp. hold
Inspiratory hold is activated by pressing the Insp. hold key. The maximum time is
30 seconds. The inspiratory and expiratory valves close after expiration. This
function can provide an exact measurement of the end inspiratory lung pressure.
It can be used during x-ray or to determine Plateau pressure, or static compliance
calculation.
Insp. hold
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
167
|
Appendix • User interface
| 13 |
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...