6.10.2 BI-VENT IN DETAIL
3
1
2
4
5
This function allows for spontaneous breathing
/ pressure supported ventilation at two
different pressure levels. These basic levels
are individually set, as well as the time in
seconds at each level.The ventilator always
tries to synchronize with the patient´s
breathing.
1.
Bi-Vent cycle; THigh + TPEEP
2.
PEEP
3.
PHigh
4.
PS above PEEP
5.
PS above PHigh
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
105
|
Ventilation, modes and functions
| 6 |