Airway Pressure
Range: -40 - 160 cmH
2
O
Inaccuracy:
±
(1 cmH
2
O + 5 % of actual
value)
Supply Pressure
Range: 0 - 7 bar
Inaccuracy:
±
5% of read value
Filtering
Pressure waveform: Low pass filtered (time
constant 15 ms)
The measured and calculated values displayed
or used for control have in some cases been
subjected to filtering and smoothing
techniques. This is done to capture the
important patterns in the data while excluding
noise and make the data shown clinically
relevant. These techniques are part of the
inaccuracy specified in the technical data.
150
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 11 |
Technical data
|
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...