6.8.2 SPONTANEOUS BREATHING/CPAP
IN DETAIL
True spontaneous breathing will occur:
- In Pressure Support when the inspiratory
pressure level is set to zero
Inspiration starts upon patient effort.
Expiration starts:
When the inspiratory flow decreases below
a preset fraction of the inspiratory peak flow
(Inspiratory cycle-off)
If the upper pressure limit is exceeded.
Maximum time for inspiration is exceeded.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
93
|
Ventilation, modes and functions