4.2.4 PATIENT CIRCUIT TEST
The test measures the compliance and
resistance of the patient breathing system.
The patient circuit resistance is automatically
measured to determine if the ventilator
maintains the specified accuracy with the
connected breathing circuit.
1.
Connect the complete breathing system
to be used on the patient. If an active
humidifier is used, it must be filled with
water.
S
VX-155a_XX
Important: When blocking the end of
the complete breathing system, make
sure there is no leakage. Leakage will
affect the circuit compliance
compensation calculation.
2.
Block the end of the complete breathing
system and follow the on-screen
instructions. The circuit compliance and
resistance are automatically measured.
Go to Compensate for Circuit compliance
S
VX-155b_XX
3.
Unblock the end of the complete
breathing system and follow the on-screen
instructions.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
45
|
Operation overview
| 4 |
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...