Always set the Apnea time appropriate to the
individual patient situation. If the apnea alarm
limit is reached the ventilator will automatically
switch to backup ventilation.
It is also very important to set lower and upper
alarm limit for expired Minute Volume.
It is important to set the Inspiratory rise time
to a comfortable value for the patient. In
Pressure Support the Inspiratory rise time
should normally be increased.
70%
Inspiratory Cycle-off is important for the
patient’s comfort and ventilator
synchronization with the patient. Inspiratory
Cycle-off is the point when inspiration
switches to expiration. E.g. for a patient with
expiratory resistance the inspiratory Cycle-off
should be set to a high value to guarantee
enough time for expiration.
Note: It is important to monitor the
corresponding Tidal Volume levels.
90
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
| 6 |
Ventilation, modes and functions
|
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...