11.8 BREATHING PARAMETERS: DEFAULT VALUES & ALLOWED
SETTINGS (STANDARD CONFIGURATION)
Setting range
Factory set default
Parameter
-
2
Bias flow (l/min)
1 - 15
4
Breath cycle time, SIMV (s)
4 - 100
15
CMV frequency (b/min)
ON/ OFF
OFF
Compensate for compliance
0 - 100%
50%
Flow trig sensitivity level (fraction of bias flow)
1:10 - 4:1
1:2
I:E ratio
1:10 - 4:1
1:2
I:E ratio in backup
1 - 70
30
Inspiratory cycle-off (% of peak flow)
10 - 70
50
Inspiratory cycle-off (% of peak flow) in NIV
0 - 20
5
Inspiratory rise time (%)
0 - 0.4
0.15
Inspiratory rise time (s)
0 - 0.4
0.2
Inspiratory rise time (s) in NIV
-
3.3
Maximum inspiratory flow (l/s)
-
120
Maximum permitted absolute pressure (cmH
2
O)
-
32
Maximum permitted absolute pressure in NIV
(cmH
2
O)
0.5 - 60
7.5
Minute Volume (l/min)
-
VC
Mode (Invasive ventilation)
-
PS
Mode (Non invasive ventilation)
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
151
|
Technical data
| 11 |
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...