11.3 STANDARD CONDITION
SPECIFICATION
Inaccuracy ranges in this document assume
the following standard conditions and the
worst case, i.e. all errors are summarized
positive. Statistically 95% of all values will be
within 2/3 of the given error.
Ambient pressure: 101.3 kPa
Room temperature: 20 ºC
Inlet pressure: 4.3 kPa x 100
Pre-use check performed on a warmed up
ventilator
Pre-use check performed with 99 %
oxygen content in O
2
supply.
Pre-use check performed with 21 ±0.5%
oxygen content in air supply.
Patient circuit compliance compensation is
activated.
Default settings unless otherwise specified
All measured, preset and indicated flows
and volumes are referenced to BTPS, see
chapter Definitions on page 163.
All measured, inlet gas pressures and flows
are referenced to STPD, see chapter
Humidifier water level is between 50 % and
100 % of the maximum level.
Set I:E is less than 1:1.
Set ventilatory frequency is less than or
equal to 100 breaths/minute.
Inaccuracy does not include measurement
uncertainty. Values including measurement
uncertainty are specified in the service
documentation available from MAQUET.
Constant leakage fraction below 30 % in
NIV modes.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
147
|
Technical data
| 11 |
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...