3.3.2 STATUS MESSAGES
Explanation
Message (message type)
Power failure.
Technical error no. 1 - 6, 29, 10001 (technical error)
There is a problem with the battery modules. One
or more battery modules must be replaced.
Check battery status (Text Message)
AC power is off line due to a power failure or
disconnection.
Battery operation (Medium Priority Alarm)
Less than 10 minutes left of battery operation.
Limited battery capacity (High Priority Alarm)
Less than 3 minutes left of battery operation.
No battery capacity (High Priority Alarm)
Battery voltage too low. Cannot guarantee
continued ventilator operation.
Low battery voltage (High Priority Alarm)
3.3.3 AC POWER FAILURE
In the event of an AC power failure or
disconnection, the ventilator switches to
battery operation and activates medium
priority alarms, see Table above in Status
Messges for details.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
39
|
Power supply
| 3 |
Содержание Maquet SERVO-s VENTILATOR SYSTEM V7.0
Страница 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Страница 2: ......
Страница 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Страница 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Страница 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Страница 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Страница 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Страница 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Страница 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Страница 177: ...14 CERTIFICATES TABLE OF CONTENTS 178 China 14 1 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 177 Certificates 14 ...
Страница 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Страница 183: ...x ...
Страница 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...