Chapter 5: Taking Exposures
Definium AMX 700 X-Ray System
5161515-1EN
Rev. 6 (10 February 2008)
5-4
Operator Manual
© 2008 General Electric Company. All rights reserved.
Table 5-1
Maximum kVp and mAs Ratings
It is important to note that the kVp setting takes precedence over your mAs setting. That is, if your
technique would generate too much heat for the x-ray tube anode, the mAs setting will
automatically decrease to permit the selected kVp.
– For instance, if kVp is set at 110 then your maximum mAs will not exceed 200.
NOTE:
The maximum allowable kVp and mAs ratings stated within this operation manual and as allowed
by the equipment, may appear to exceed the tube ratings. The reason for this apparent
discrepancy is that the tube rating charts in Product Data Sheet 2236721-100 are based upon a
25% heated tube assembly (75% heat units remaining), while the actual maximum allowable kVp
and mAs ratings are based upon a cold tube assembly. The Definium AMX 700 generator contains
a proprietary heat loading algorithm which protects the tube assembly for all heat loading
conditions.
kVp
mAs
110 to 125
200
95 to 105
250
50 to 90
320