Chapter A4: Image Acquisition
Definium AMX 700 X-Ray System
5161515-1EN
Rev. 6 (10 February 2008)
A4-6
Operator Manual
© 2008 General Electric Company. All rights reserved.
Re-start Completed or Discontinued Exams
Procedures with a Status of “Completed” or “Discontinued” cannot be re-opened. However, you may
create a new exam or append the existing exam.
1. Select the procedure(s) from the Worklist.
2. Press [PATIENT INFORMATION] to verify the patient, if necessary.
3. Press [RESUME EXAM].
A message appears: “The selected exam has been Discontinued/Completed and cannot be
started/resumed. Would you like to create a new exam for this patient or append to the existing
exam?”
Figure A4-2
Start a discontinued or completed exam message
4. Press the button of the action to perform.
[APPEND EXAM] creates a new series within the existing exam and opens the Select Protocols
screen.
[QAP]
When the yellow alert icon is present, indicates that QAP should be
performed to ensure continued image quality.
NOTE:
QAP cannot be performed when the Image Acquisition or
Image Viewer screens are displayed. End the exam before
attempting to perform QAP.
Chapter A8: Quality Assurance Process and Maintenance
for more information.
[EXPOSE HOLD]
Appears when there is some condition that prevents an x-ray from
being taken.
Press the button to view a list of all errors and interlocks that are
preventing the exposure. The items are removed from the list as
they are corrected. The button disappears when all errors and
interlocks are corrected.