Software Start Up and Shutdown
This section describes the procedure for starting up and shutting down the GE Portable DR Detector
software only.
Start Up
Chapter 7: Maintenance and Service-Periodic Maintenance by Service Personnel
(p. 7-1) for
important information about the circuit breaker.
1. Turn the key switch to the ON position.
The unit performs diagnostics (self-tests to make sure everything is working correctly). This
process takes a few seconds.
2. When the technique and battery status displays light up, the unit is ready for operation. TESTING
COMPLETE will also briefly appear on the message display.
The unit may take 3 minutes for the digital detector software to boot up. You may drive the unit
before the software finishes booting, but you cannot take acquisitions until the software is fully
booted.
3. If SYSTEM -- XX FAILED is displayed, turn the unit OFF then ON again. If SYSTEM -- XX FAILED displays
again, write down the alphanumeric code (XX represents the code) and contact your service
representative.
Shutdown
1. Close all current exams.
2. Press the [UTILITY] button at the top of the Worklist screen.
Figure A2-3
Utility button
3. Select
System
on the Utilities screen.
Figure A2-4
Utilities – System screen