Chapter 6: Charging the Batteries
Definium AMX 700 X-Ray System
5161515-1EN
Rev. 6 (10 February 2008)
6-6
Operator Manual
© 2008 General Electric Company. All rights reserved.
Main Power Circuit Breaker
The main power circuit breaker is located on the front right side of the unit (Figure 6-5).
Figure 6-5
Main power circuit breaker
If the battery charge is too low, or in the event of an overload in the high-voltage system, the circuit
breaker will trip (interrupt power to the x-ray unit).
Should the circuit breaker trip, the control panel lights will go out and you will be pre vented from
operating the unit. To resume operation, flip the circuit breaker up to the ON position. If necessary,
recharge the unit or (if the battery is adequately charged) try the procedure again.
If the battery is within operating condition (as indicated by the battery status and message displays) and
power interruptions are frequent, there may be a fault in the high voltage system. If a fault is indicated,
have the unit checked by a qualified service representative.
If the unit does not turn off when the key switch is turned off, isolate the unit to prevent unauthorized
operation, use the circuit breaker to remove power from the unit, and have the unit checked by a
qualified service representative.