7-32
170 Series Monitor
Revision C
2000947-004
Serviceable Assemblies:
Nautilus Transducer Reassembly
Nautilus Transducer Reassembly
CAUTIONS
VISIBLE INSPECTION—Ensure the base sealing groove, flat-
seal o-ring, sealing surface, and sealing screws are free of visible
surface defects, dust, dirt, and foreign material.
SINGLE-USE COMPONENTS—Do
not
reuse case top, flat-seal
o-ring, sealing screws, or screw caps. Discard these items
each
time a transducer is opened.
SEALS—Wetting of seals is
not
required.
1. Install new flat-seal o-ring.
2. Install case top and tighten five sealing screws. Hand tighten each screw a small
amount going in a clockwise direction, skipping every other screw. Repeat until
all screws are torqued to 48 in-oz.
3. Apply one drop of Loctite adhesive
1
to the perimeter of the hole for the flat
). Insert flat screw cap and press into place.
4. Apply one drop of Loctite adhesive
to the perimeter of the hole for a curved
screw cap. Insert the curved screw cap being careful to align it according to the
curvature of the case top; the small burr on the screw cap side should face the
outside edge of the transducer. (See
.) Press the
screw cap firmly into place. Repeat for the remaining three curved screw caps.
5. For button-style transducers only, install and torque belt knob to
48 in-oz.
6. Refer to
“Testing a Repaired Transducer (TOCO or US)”
1
Using Loctite Adhesive #454 is recommended in order to secure the screw caps.
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......