Revision C
170 Series Monitor
7-5
2000947-004
Serviceable Assemblies:
Transducer Plug Replacement Kits
Tocotransducer Cable Preparation
Refer to
and
while following these steps:
1. Cut #24 AWG green wire (not supplied) to 1-
1
⁄
2
inches long and strip each end
1
⁄
8
inch. This will be used as a jumper wire.
2. Cut #24 AWG green wire to 2 inches long. Strip one end
3
⁄
4
inch.
3. Route the transducer cable through the gripper. Slide the gripper, followed by
the retaining ring, down the cable until they are out of the way.
4.
For Nautilus Tocotransducers only
: Slide Oetiker clamp over end of cable. This
clamp is not necessary for Trimline Tocotransducers.
5. Score around grey cable jacket 1 inch. Peel and remove jacket. Trim filler flush
to jacket.
6.
For Nautilus Tocotransducers only:
Cut vent tubing flush with jacket being
careful not to restrict air flow. The vent tubing is not present in Trimline cables.
7. Pull back the braided shield and twist into pigtail.
8. Solder the 2-inch prepared green wire to the shield, wrapping the
3
⁄
4
-inch strip on to the shield. Cut wire flush with other conductors of the cable.
9. Install shrink tubing over the solder joint of the shield.
10. Strip all conductors
1
⁄
8
inch. Twist and tin conductors.
11. Locate the grounding contact pin (longer pin) in the kit. Solder this pin to the
shield wire.
12. Solder a short pin to the yellow wire
and
one end of the green jumper wire.
Both wires share this pin.
13. Solder a short pin to the free end of the green jumper wire.
14. Solder a short pin to each of the other conductors.
15.
For Nautilus Tocotransducers only:
Crimp Oetiker clamp over cable jacket
1
⁄
8
inch from end.
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......