Serviceable Assemblies:
Nautilus Ultrasound Transducer Top Cover Replacement
Nautilus Ultrasound Transducer Top Cover
Replacement
In order to replace the top cover of a Nautilus Ultrasound Transducer in the field,
you will require one of two kits. Refer to
and ensure you have the correct
kit. Unpack the kit and ensure it contains the items listed in
. If
something is missing, contact your GE Service Representative immediately. The
text of the instruction sheet is repeated here for informational purposes.
.
Equipment Required
You will need the following equipment:
Phillips Head Torque Driver
Loctite Adhesive #454 (recommended)
Anti-Static Wristband
Static-Free Work Surface
Table 7-11. Kit Summary
Nautilus Ultrasound Transducer
Type and Cat. No. (REF)
Cable Kit
Cat. No. (REF)
Loop-Style Ultrasound Transducer, 5700 KAX/LAX/MAX
2011679-001
Button-Style Ultrasound Transducer, 5700GAX/HAX/JAX
2011680-001
Table 7-12. Packing List
Item
Qty
Transducer Case Top
1
Flat-Seal O-Ring
1
Sealing Screw, #4-40 x 5/16 in
6
Screw Cap, Curved
4
Screw Cap, Flat
1
Belt Knob
a
a
The belt knob is included in button-style
kits 2011680-001 only.
Screw Cap Extraction Tool
1
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......