Revision C
170 Series Monitor
6-25
2000947-004
Functional Checkout Procedure:
Dual Heart Rate Test (Non-Pattern)
Table 6-6. Model 325 Simulator Settings for FECG/US Test
Section
Switch
Setting
FECG/MECG
Main
RATE
Rate
120 BPM
Mode
FECG
QRS Amplitude
50 µV
QRS Polarity
+
ULTRASOUND/FMD
Mode
US
Level
MED
Rate
RAMP
General
Pattern Memory
OFF