6-2
170 Series Monitor
Revision C
2000947-004
Functional Checkout Procedure:
Equipment Required
Equipment Required
provides a complete listing of equipment required for a functional
checkout procedure.
Table 6-1. Equipment Required
Test
Equipment
FECG
Model 325 Simulator (14203A/B)
Monitor Interconnect Cable (14212A/B)
Legplate
Legplate for Qwik Connect Plus with round connector
(1590AAO)
US/FMD
Model 325 Simulator (14203A/B)
Monitor Interconnect Cable (14212A/B)
US Transducer
Nautilus Ultrasound Transducer
(5700GAX–5700MAX Series)
TOCO/IUP
Model 325 Simulator (14203A/B)
Monitor Interconnect Cable (14212A/B)
Tocotransducer
Nautilus Tocotransducer
(2264GAX–2264MAX Series)
Strain Gauge Transducer
Reusable Strain Gauge with holder (4007BAX); or
Reusable Strain Gauge without holder (4007LAX); or
Disposable Strain Gauge (4009AAX)
NOTE: These items are no longer sold by GE. However,
transducers in the field are compatible with the 170
Series Monitor.
Pattern Memory
Model 325 Simulator (14203A/B)
Monitor Interconnect Cable (14212A/B)
Dual Heart Rate (FECG/US)
Model 325 Simulator (14203A/B)
Monitor Interconnect Cable (14212A/B)
Dual Heart Rate (US/US2)
Two Model 325 Simulators (14203A/B) with two sets
of Monitor Interconnect Cables (14212A/B); or
Two Nautilus Ultrasound Transducers
5700GAX–5700MAX Series)
Alarms
Model 325 Simulator (14203A/B)
Monitor Interconnect Cable (14212A/B)
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......