7-48
170 Series Monitor
Revision C
2000947-004
Serviceable Assemblies:
Servicing the Recorder
18. Replace the microswitch and secure both screws. Make sure the microswitch
clicks when the drawer is closed.
19. Replace the main board mounting plate, pulling the plate forward, parallel with
the roller while tightening the screws.
20. Return the printhead assembly to its original position as noted in
step 4.
21. Replace the main board, re-connect all cables, and replace the monitor top
cover.
22. Check printing and re-align the printhead as necessary. (See
on
.)
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......