8-4
170 Series Monitor
Revision C
2000947-004
Peripheral Devices:
RS-232 Connectors
RS-232 Connectors
Two RS-232C serial interface receptacles (RJ-45) allow connecting the 170 Series
Monitor to the following devices:
Critikon Model 1846/1846SX Blood Pressure Monitor
Quantitative Sentinel/Perinatal System
Nellcor N-400 Fetal Pulse Oximeter
115-Compatible Device
Table 8-1 lists the factory default settings for each port. Refer to
Customizing the
Monitor
on page 4-10 for information on using the service setup mode to change
these settings.
Baud Rate
Select a baud rate that is compatible with the external device. The available settings:
600, 1200, 2400, 4800, 9600, 19,200, and 384,000 bps.
Mode
Select a communications mode compatible with the external device.
HP or HP with Notes
Select HP or HP w/notes for connecting to a Marquette Quantitative/Perinatal
System or another central system that uses the Hewlett Packard communications
protocol. HP mode does not print central station annotations on the fetal monitor
strip chart, whereas HP w/notes does.
Ext. BP
Select ext. BP for connecting to a Critikon Model 1846/1846SX Blood Pressure
Monitor.
Table 8-1. Communication Default Settings
RS-232 Port
Intended Device
Mode
Baud Rate
1
QS System
HP
1200
2
Critikon Model 1846/1846SX
ext. BP
600
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......