5-8
170 Series Monitor
Revision C
2000947-004
Theory of Operation:
Functional Overview
* Active low.
Figure 5-5. US Connector(s) (facing the front panel from the outside)
Table 5-9. Ultrasound Connector(s)
Pin #
Signal Name
Signal Description
1
NC
No Connection
2
NC
No Connection
3
GND
Chassis Ground
4
XMIT/RCV SHIELD
Shield for Transmit/Receive
5
XMIT/RCV
Transmit/Receive
6
NC
No Connection
7
NC
No Connection
8
NC
No Connection
9
GND
Chassis Ground
10
NC
No Connection
11
U/S ENABLE*
Enable for Ultrasound Channel
12
GND
Chassis Ground
1
2
3
4
5
6
7 12
9
8
10
11
2
3
4
5
6
8
7
9
10
11
12
1
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......