6-22
170 Series Monitor
Revision C
2000947-004
Functional Checkout Procedure:
Strain Gauge Transducer Test
Strain Gauge Transducer Test
1. Inspect a strain gauge as follows:
Unscrew the plastic dome from the transducer and check for any cracks or
contaminants on the transducer.
Visibly assess the condition of the cable, strain relief, and the connector
pins.
2. Connect the strain gauge to the
UA
input receptacle on the front panel of the
monitor. Verify the following on the monitor:
The UA value may read negative numbers indicating baseline pressure is
off scale. In this case, the message
BASELINE PRESSURE OFF SCALE
prints on
the strip chart paper.
After approximately 20 seconds, the message
IUP
prints on the center
margin of the strip chart paper.
3. Select the monitor’s
UA Reference
button and verify the following on the
monitor:
The UA value is 0 mmHg.
The UA trend is a continuous line at 0 mmHg on the strip chart paper.
The message
UA REF
prints on the strip chart paper.
4. Apply gentle pressure on the strain gauge diaphragm and verify that the UA
value and trend respond to the input. Increasing force should produce an
increasing value and vice versa.
5. Disconnect the strain gauge from the front panel of the monitor. Verify the
following on the monitor:
The UA value is blank.
The UA trend stops plotting on the strip chart paper.
After approximately 20 seconds, the message
UA INOP
prints on the strip
chart paper.
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......