Revision C
170 Series Monitor
5-31
2000947-004
Theory of Operation:
Main Board Theory of Operation
Figure 5-18. Audio Module Block Diagram
Ultrasound Envelopes
The outputs of the main filters are also input to U32 and U35. The input to U35 is
through high-pass filters at 63 Hz. The high pass components for Channel 1 and
Channel 2 are C154/R338 and C159/R357, respectively. The output of the main
filters are also input to U32; this integrated circuit, in conjunction with U35, forms
an AGC (automatic gain control) circuit for both ultrasound channels.
Each of these AGC circuit has a maximum gain of 200 with a dynamic range of 46
dB. Each of the AGC outputs are high-pass filtered by
C169/R343 at 90 Hz with a gain of one, then input to a half-wave detector U36 (pin
7) with a gain of 1.47. The output of the detector circuits are low-pass filtered by
U36 (pin 1)at 20 Hz at gain of ten.
These envelopes are then input to the analog-to-digital converter U38. Other inputs
to the analog-to-digital converter are the telemetry TOCO signal. This signal is
scaled and offset to provide –100 to +400 relative UA units.
Digital
Potentiometers
for Volume
Control (x3)
Ultrasound
Audio
Doubler
Bridge
Amplifier
Summing
Amplifier
Summing
Amplifier
Tone Bit
Channel 1
US Audio
Channel 2
US Audio
Serial Clock and Data
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......