Functional Checkout Procedure:
Connecting the Simulator
Connecting the Simulator
This part of the procedure prepares the simulator for use.
NOTE:
You must use a Model 325 Simulator for the functional checkout
procedure. (Monitors in the 170 Series do not work with Model 305
Simulators.)
1. Ensure the Model 325
Power
switch is in the
off
position.
2. Connect the Model 325 Simulator’s power cord to the power receptacle on the
rear panel of the simulator; plug the other end of the power cord into a properly
grounded wall outlet of appropriate voltage.
3. Ensure the 170 Series Monitor is turned off.
4. Connect the simulator interconnect cable’s 50-pin end to the simulator’s
Fetal
Monitor
connector.
5. Connect the sub-cables of the other end of the simulator interconnect cable into
the color-coded connectors on the monitor/adapter:
US
and
UA
(as required by
the procedure).
6. Turn
on
the Model 325 Simulator. Verify that the green
Power
indicator
illuminates.
7. Turn
on
the 170 Series Monitor.
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......