7-24
170 Series Monitor
Revision C
2000947-004
Serviceable Assemblies:
Tocotransducer Calibration
Calibration for Cat. No. (REF) 2264 AAX/BAX/CAX/DAX/EAX/FAX
during calibration.
1. Set the Gain (R15) and Offset (R5) potentiometers fully counterclockwise.
2. Set the Offset Trim (R6) potentiometer to its mid-position—approximately
six turns from fully counterclockwise.
3. Connect the positive lead of the digital voltmeter to TP2 on the Main Board;
connect the negative lead to TP9.
4. Turn on the monitor.
5. Plug the CMR Test Jack into the monitor’s
UA
connector.
6. Measure the voltage across the test points and record the value.
7. Replace the CMR Test Jack with the transducer to be calibrated.
8. Place the tocotransducer on a flat surface so that the button is face up, then
center the 52.5 gram weight on the pressure sensitive button.
9. Adjust the Offset (R5) and Offset Trim (R6) potentiometers so that the
voltage matches the reading recorded in step 6
±
0.04 V. This voltage is
referred to as the “weight on” voltage.
The Offset (R5) potentiometer provides course adjustment.
The Offset Trim (R6) potentiometer provides fine adjustment.
10. Remove the weight and observe the voltage reading; this voltage is referred
to as the “weight off” voltage. A correctly adjusted transducer will show a
voltage change of +0.115 V
±
0.015 V. If the change is less, adjust the Gain
(R15) potentiometer counterclockwise to increase the gain. If the change is
more, adjust the Gain (R15) potentiometer clockwise to decrease the gain.
11. Repeat step 8 through step step 10 until the change in voltage between the
“weight on” and “weight off” conditions falls into specification.
12. Torque seal all potentiometers.
13. Follow the instructions for
“Nautilus Transducer Reassembly”
on
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......