Revision C
170 Series Monitor
6-27
2000947-004
Functional Checkout Procedure:
Dual Heart Rate Test (Non-Pattern)
Dual Ultrasound Modes
As stated previously, the dual ultrasound mode of the 170 Series Monitor cannot be
tested unless two Model 325 Simulators are used or two Nautilus ultrasound
transducers.
Do not
attempt to test dual ultrasound using one Model 325 Simulator
and one ultrasound transducer. This procedure details using two transducers since it
is more practical for a test site.
1. If not already on, press the monitor’s
Record
pushbutton.
2. Plug one ultrasound transducer into the monitor’s
US
input receptacle and the
other into the monitor’s
US2
receptacle. Verify the following on the monitor:
The FHR1 value shows three steady dashes “– – –.”
The FHR2 value shows three steady dashes “– – –.”
The recorder prints the messages
US
and
US2
on the center margin of the
strip chart paper.
3. Use your finger to rub the face of the ultrasound transducer connected to the
monitor’s
US
input receptacle; try to maintain a steady rate and verify the
following on the monitor:
The FHR1 value responds to the rubbing.
The FHR1 heartbeat indicator (
) responds to the input.
The recorder prints the heart rate tracing corresponding to the rate and the
trace is plain black (
).
4. Use your finger to rub the face of the ultrasound transducer connected to the
monitor’s
US2
input receptacle; try to maintain a steady rate and verify the
following on the monitor:
The FHR2 value responds to the rubbing.
The FHR2 heartbeat indicator (
) responds to the input.
The recorder prints the heart rate tracing corresponding to the rate and the
trace is bold black (
).
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......