9-4
170 Series Monitor
Revision C
2000947-004
Maintenance:
Cleaning
Cleaning the UA Strain Gauge
1. Remove the plastic dome.
2. If desired, wash the transducer with sterile water or saline solution.
3. Carefully clean the diaphragm seal with a cotton swab to remove deposits.
Avoid excessive pressure since this may damage the diaphragm. If there are
excessive stains on the diaphragm or sides of the transducer, remove with a
cotton swab and solvents of increasing strength. Do not use pumice, Ajax, Bon
Ami or other abrasives.
4. After cleaning, rinse the transducer thoroughly in distilled water and replace the
dome loosely.
5. Dry the transducer with sterile gauze.
CAUTIONS
AUTOCLAVE—Do not autoclave pressure transducer.
IMMERSION—Do not immerse any part of the electrical
connector of the transducer in the cleaning solution at any time.
Examine the outer sheath of the cable for perforations. If the outer
covering is damaged in any way, do not immerse the cable in the
cleaning solution; this may result in moisture entering the
transducer case, which is vented through the cable.
WARNING
LIQUIDS—If liquids enter the electrical connector, check the
resistance between the electrical element and the transducer case.
A resistance level of greater than 10 M¾ ensures that the leakage
current is within acceptable levels for safe use on patients.
6. Leave transparent dome attached to the transducer during storage, but slacken the
locking ring at least one quarter of a turn.
CAUTION
STERILIZATION—Prior to patient use, ensure the dome is sterile.
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......