Functional Checkout Procedure:
Alarm Test
26. Use the simulator’s
Manual Adjustment
knob to increase the US signal to 100
bpm. Verify the following on the monitor:
The alarm tone continues sounding.
The Alarm indicator stops continues flashing.
The affected heart rate numerics continue flashing.
27. Press the monitor’s front panel
Alarm Silence
pushbutton. Verify that the
visual
and
audio alarm indications disappear.
28. This concludes the functional checkout procedure.
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......