9-14
170 Series Monitor
Revision C
2000947-004
Maintenance:
Preventative Maintenance Inspection
Patient-to-Line Leakage for TOCO
1. Connect a tocotransducer test body to the monitor’s front panel
UA
input.
2. Configure a leakage tester to perform a line leakage test.
3. Turn ON the monitor using the monitor’s
On/Standby
button.
4. Verify the following:
<50
µ
A
(132–253 VAC, or at the line voltage at your fa10%)
5. Record the patient-to-line leakage for the following conditions:
Table 9-7. Patient-to-Line Leakage Conditions for TOCO
Neutral
Ground
Power (Polarity)
__________ Closed
__________ Closed
__________ Normal
__________ Closed
__________ Closed
__________ Reversed
__________ Open
__________ Closed
__________ Normal
__________ Open
__________ Closed
__________ Reversed
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......