7-16
170 Series Monitor
Revision C
2000947-004
Serviceable Assemblies:
Nautilus Transducer Cable Replacement
Equipment Required
You will need the following equipment:
Phillips Head Torque Driver
Screw Cap Extraction Tool (cat. no. 2004084-001)
Soldering Iron and Solder (ultrasound transducers only)
Loctite Adhesive #454 (recommended)
Anti-Static Wristband
Static-Free Work Surface
Reference Figures
Refer to
and
while using the cable replacement kit.
Figure 7-5. Sealing Screw Caps
Figure 7-6. Nautilus Transducers—Cover Removed
Flat Screw Cap
Button-Style
Loop-Style
Tocotransducer
2264 AAX/BAX/CAX/DAX/EAX/FAX
Cable Nut
Set Screw
Tocotransducer
2264 GAX/HAX/JAX/KAX/LAX/MAX
Self-Tapping
Screws
Green Wire
(Shield)
Clear Wire
(Center Conductor)
Green Wire
(Ground)
Cable Nut
Set Screw
Ultrasound Transducer
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......