Number of inlets
Data
Parameter
7, expandable to 14
Inlet A
7, expandable to 14
Inlet B
7, expandable to 14
Sample inlet
4
Quaternary inlet
Pressure monitors
Data
Parameter
4
Number of sensors
•
The System pressure monitor is located after the System
pumps.
•
The Pre-column pressure monitor and the Post-column
pressure monitor are integrated in the Column valve V9-
C or V9H-C.
•
The Sample pressure monitor is located after the Sample
pump.
Placement of sensors
ÄKTA avant 25: 0 to 20 MPa (0 to 200 bar)
Range
ÄKTA avant 150: 0 to 5 MPa (0 to 50 bar)
ÄKTA avant 25: ± 0.02 MPa or ± 2% whichever is greater
Accuracy
ÄKTA avant 150: ± 0.015 MPa or ± 1.5% whichever is greater
UV monitors
Description
Item
Up to 2
Number of monitors
U9-M: 190 to 700 nm in steps of 1 nm, up to 3 wavelengths
Wavelength range
U9-L: 280 nm
-6 to 6 AU
Absorbance range
ÄKTA avant User Manual 29035184 AE
427
9 Reference information
9.2 Module specifications