Possible cause and action
Problem
•
One or some of the minor modules use too much
current. Minor modules include all modules except
the Fraction collector, the UV monitor and the
pumps.
The current is cut off by a temperature sensitive
component. The instrument can be restarted when
the temperature has decreased. If persistent, generate
a System error report and contact Service.
One or more modules
are automatically
turned off
•
The network cable is not connected
Connect the cable, see UNICORN Administration and
Technical Manual.
The instrument cannot
be seen in UNICORN
•
The cable between a module and the ICU is not
connected.
Remove valves and check that cables are connected.
•
One or more jumpers are loose or missing.
Check the connections on the back of the instrument.
Empty positions must have connected jumpers. See
the illustration below.
One or more modules
are not found by the
instrument
408
ÄKTA avant User Manual 29035184 AE
8 Troubleshooting
8.7 Troubleshooting: Other components