UV Monitor U9-L
Action
Description
Error
code
See UV monitor U9-L, on page 368.
High temperature
51
Contact Service.
Low lamp intensity
52, 55
Autozero requested when AU value is
larger than 2.
Autozero out of range
54
No light through flow cell. Check solution
absorption and that the cell is fitted cor-
rectly.
Too low light intensity, S
channel
58
Restart the instrument. If recurrent please
contact Service.
Internal instrument error
59, 60
Restart the instrument. If recurrent please
contact Service.
Measurement error
61
Fraction Collector F9-R
Action
Description
Error
code
See Fraction Collector F9-R, on page 394
High temperature
20
Clean the sensor and remove air bubbles
in the flow path.
DropSync warning
21
Check that the tube sensor is adjusted
properly.
Tube sensor error
22
Increase the fraction size. Fraction size
must be greater than 1/10 of the delay
volume.
Delay queue full
23
Restart the instrument. If recurrent please
contact Service.
Internal instrument error
24, 26
Increase the fraction size or lower the
flow rate.
Too fast tube change
25, 28
Clean the drop sensor
DropSync error
27
ÄKTA avant User Manual 29035184 AE
419
8 Troubleshooting
8.9 Error codes