Action
Step
Unscrew the knurled wheel at the bottom of the UV monitor. Press the wheel
upwards to release the flow cell.
3
Pull the flow cell upwards out of the monitor. Hold the flow cell by the top
part with the O-ring: do not touch the optical surfaces of the flow cell.
4
Note:
Make sure that the flow cell does not come into contact with any liquid, and
that no liquid enters the monitor.
Insert a new flow cell into the monitor.
5
Tighten the knurled wheel firmly.
6
Connect the tubing to the new flow cell.
7
Switch on the instrument and log on to UNICORN.
8
Update the UV flow cell path length in the Calibrate dialog box, in the
System Control module.
9
346
ÄKTA avant User Manual 29035184 AE
7 Maintenance
7.8 Replacement procedures
7.8.5 Replace UV monitor U9-L flow cell