7.8.1
Replace tubing and connectors
Maintenance interval
Replace tubing and connectors when required, for example when a tubing has clogged
or has been bent so that the flow is stopped.
Required material
The following materials are required:
•
Tubing and connectors
•
Tubing cutter
•
Fingertight wrench
Instruction
Follow the instruction below to replace tubing and connectors.
Action
Step
Make sure that no run is in progress on the instrument.
1
Unscrew the connectors, and disconnect the tubing.
2
If the tubing has labels, remove the labels to be used with the new tubing
later. Discard the tubing and connectors.
3
336
ÄKTA avant User Manual 29035184 AE
7 Maintenance
7.8 Replacement procedures
7.8.1 Replace tubing and connectors
Содержание AKTA avant