
Action
Step
In the Calibration dialog, in the Monitor to calibrate list, select the pressure
monitor to calibrate.
2
Click Reset pressure.
3
Result: The atmospheric pressure is defined as zero. Date and time of the
most recent calibration, and the current pressure value are displayed.
Reconnect the tubing to the pressure monitor.
4
324
ÄKTA avant User Manual 29035184 AE
7 Maintenance
7.7 Calibration procedures
7.7.2 Calibrate the pressure monitors
Содержание AKTA avant
Страница 1: ...ÄKTA avant User Manual ...
Страница 2: ...Page intentionally left blank ...
Страница 536: ...Page intentionally left blank ...
Страница 537: ...Page intentionally left blank ...