Overview of the I/O-box
Function of the I/O-box
The I/O-box E9 is used to interface other equipment in order to measure parameters
such as refractive index, light scattering and fluorescence. See Requirements on connected
equipment, on page 150 for information on requirements of the equipment that can be
connected to ÄKTA avant. The I/O-box can control external equipment by a digital output
signal, as well as detecting the state of them by digital inputs. It is also possible to send
out internal detector signals to external equipment.
Using two I/O-boxes
It is possible to install up to two I/O-boxes when using ÄKTA avant. If two I/O-boxes are
to be used, the second I/O-box has to be configured as I/O-box E9, 2nd. The configuration
is defined by the Node ID of the I/O-box. The Node ID is set by positioning the arrow of
two rotating switches at the back of the I/O-box, see Connectors, on page 145 and
Section 9.15 Check and change the Node ID of a module, on page 515.
Location
The most common placement of the I/O-box, with the adhesive feet attached, is on the
laboratory bench on the right-hand side of the instrument main wet surface, close to
the UniNet ports. It is an advantage to have the external equipment that is going to be
connected close to the I/O-box on the same side of the instrument. The I/O box can also
be placed on a Rail extension rod, see Rail extension, on page 51.
144
ÄKTA avant User Manual 29035184 AE
4 Optional instrument modules
4.12 I/O-box
4.12.1 Overview of the I/O-box
Содержание AKTA avant
Страница 1: ...ÄKTA avant User Manual ...
Страница 2: ...Page intentionally left blank ...
Страница 536: ...Page intentionally left blank ...
Страница 537: ...Page intentionally left blank ...